Free YouTube views likes and subscribers? Easily!
Get Free YouTube Subscribers, Views and Likes

How to use ModBusPoll On a TacticalFlow MAG meter

Follow
Dave Korpi

How to use ModBusPoll On a TacticalFlow MAG meter

You will want to make sure your MAG meters Address is set to agree with your ModBus Poll and the BAUD rate is set to the value you wish. Default on the MAG meter is Slave ID 1 and 9600 BAUD, No Parity, 1 stop and 8 bits. The very last menu on the user screen is the parity and stop.

We demonstrate reading an easily verifiable value that is called MTP on the Meter display. To get this value to display use the Up button to scroll until you see the MTP annunciator and it will look like this:
MTP 003930% if the meter is on the bench in air. Note: the meter will NOT be harmed if run with no water.

To get into the meter to configure see our video here...
   • How to set UNITS in a MAG meter made ...  

Super simple and super fast. The password is 0321

https://www.tacticalflowmeter.com/pro...

Call (831)2448080for questions
Tactical Flow Meter MAG meter features LARGE LCD display, shows the Instantaneous Flow Rate, Total Flow, in user selectable units.

Specifications.

Electrode Material: 316L SS
Wetted Materials Teflon, 304 SS, & 316 SS
Flow Body Material 304 SS
Flow Velocity .01 15 m/s Nominal 0 3 m/s
Accuracy: +/ 0.5% of reading in water.
Repeatability of +/ 0.15% of reading
1/2" to 4" 150# ANSI flange connections, up to 6" available custom.
LCD Display
Temperature:
Gas: 14°F (10°C) to 350°F (180°C)
Ambient: 14°F (10°C) to 140°F (60°C)
Power Requirement: 24 VDC 20 Watts maximum;
Output signals: Modbus RTU, 4–20 mA, and 010 KHz for flow rate indication
Digital communications: Modbus RTU
Proof Pressure 1.6 MPa, 230 PSIG.
Display: Flow rate, Flow rate 0100%, and total flow
Wiring connection to enclosure size: M20 x 1.5
Factory Final QC Test Certificate
Integral Electronics
Electronics Enclosure: IP65

View full spec sheet here.
https://www.tacticalflowmeter.com/pro...

posted by hal3l2yn5